January 17, 2014 Visit 2
|
|
|
Maquet Servo i
I did an annual PM on the Maquet Servo i ventilator and also installed a maintenance kit. The maintenance kit contained: 2 filters for the gas modules, 2 nozzle units for the gas modules, bacteria filter for O2 cell, and a bacteria filter for inspiratory pressure transducer. The two gas modules are medical grade air and oxygen. I also performed an electrical safety check. All of the items on the electrical safety inspection were within specifications. While doing a physical inspection of the ventilator, it was noticed that the brakes were not functioning. A corrective work order was completed and new casters were ordered. I later completed another electrical safety inspection and it was within the specifications. Both ventilators were returned to service.
|