Spinchron 15R
A work order came in that this centrifuge was shaking and making a knocking sound when it was running. We verified the problem and then took the buckets and tube holders out. We checked the weight. There are 4 buckets and 2 of them weigh 215 grams and the other 2 weigh 228 grams. The positioning of the buckets was causing the centrifuge to be out of balance. We cleaned the centrifuge and put the buckets and tube holders back in their correct configuration. An electrical safety inspection was also completed.
|
|
|
The rest of the day...
The rest of the day I toured several different imaging areas. I saw imaging equipment that included x-ray, MRI, PET, CT, and Fluoroscopy. I did not get many pictures because most of the machines were larger than I could get in my screen. A few of the machines that I saw were the DRX Mobile, DRX Evolution, Light Speed CT, Millennium MG, Infinia, DRX Revolution Discovery PET, Axiom Sireskop SD, Coroskop Plus, Sireskop SX, and the Artis Zee cath lab.
|